CALibration:STRing <quoted string>
Record calibration information about your multimeter. For example,
you can store such information as the last calibration date, the next
calibration due date, the instrument serial number, or even the name
and phone number of the person to contact for a new calibration.
[Stored in non-volatile memory]
You can record information in the calibration message only from the
remote interface. However, you can read the message from either the
front-panel menu or the remote interface.
The calibration message may contain up to 40 characters. However,
the multimeter can display only 12 characters of the message on the
front panel (additional characters are truncated).
CALibration:STRing?
Query the calibration message and return a quoted string.
CALibration:VALue <
value>
Specify the value of the known calibration signal used by the calibration
procedure.
CALibration:VALue?
Query the present calibration value.