Agilent Technologies 34401A Camera Accessories User Manual


 
703 Invalid secure code
An invalid calibration security code was received when attempting to
unsecure or secure the multimeter. You must use the same security code
to unsecure the multimeter as was used to secure it, and vice versa. The
security code may contain up to 12 alphanumeric characters. The first
character must be a letter.
704 Secure code too long
A security code was received which contained more than 12 characters.
705 Cal aborted
A calibration in progress is aborted when you press any front-panel key,
send a device clear, or change the local/remote state of the multimeter.
706 Cal value out of range
The specified calibration value (CALibration:VALue) is invalid for the
present function and range.
707 Cal signal measurement out of range
The specified calibration value (CALibration:VALue) does not match
the signal applied to the multimeter.
708 Cal signal frequency out of range
The input signal frequency for an ac calibration does not match the
required input frequency for calibration.
709 No cal for this function or range
You cannot perform calibrations for ac current, period, continuity, diode,
ratio, or on the 100 M
range.
710 Full scale correction out of range
720 Cal DCV offset out of range
721 Cal DCI offset out of range
722 Cal RES offset out of range
723 Cal FRES offset out of range
724 Extended resistance self cal failed
5
Chapter 5 Error Messages
Calibration Errors