To Store Minimum and Maximum Readings
You can store the minimum and maximum readings during a series
of measurements. The following discussion shows how to read the
minimum, maximum, average, and reading count.
You can use min-max with any function except continuity or diode test.
The min-max operation is local to the selected function; when you
change functions, min-max is disabled.
After enabling min-max, you can read the stored minimum,
maximum, average, and count by pressing
Shift > (Menu Recall).
This takes you to the “
MIN–MAX” command in the MATH MENU
(only if min-max is enabled). Go down to the “parameter” level,
and then read the values by pressing
< or > .
The stored values are cleared when you turn min-max off, turn off the
power, or perform a remote interface reset.
The average is of all readings taken since min-max was enabled (not
just the average of the stored minimum and maximum). The count is
the total number of readings taken since min-max was enabled.
Enables min-max operation;
Press again to disable.
Math annunciator is on when
min-max operation is enabled.
To read the minimum, maximum, average, and count,
use the MATH menu.
2
Chapter 2 Front-Panel Menu Operation
To Store Minimum and Maximum Readings
39