Agilent Technologies 34401A Camera Accessories User Manual


 
Remote Interface Selection
The multimeter is shipped with both an GPIB (IEEE-488) interface
and an
RS-232 interface. Only one interface can be enabled at a time.
The
GPIB interface is selected when the multimeter is shipped from
the factory.
The remote interface can be set only from the front-panel.
The interface selection is stored in non-volatile memory, and does not
change when power has been off or after a remote interface reset.
If you select the GPIB interface, you must select a unique address for
the multimeter. The
GPIB address is displayed when you turn on the
multimeter.
If you select the RS-232 interface, you must set the baud rate and
parity for the multimeter. “
RS-232” is displayed when you turn on the
multimeter.
If you select the RS-232 interface and then set the GPIB address to the
talk only address (31), the multimeter will send readings over the
RS-232 interface when in the local mode.
There are certain restrictions to be aware of when you are selecting
the remote interface (see also “Programming Language Selection,” on
page 94). The only programming language supported on
RS-232 is SCPI.
Front-Panel Operation:
2: INTERFACE (I/O MENU)
See also “To Select the Remote Interface,” on page 162.
SCPI Language
3478A Language
Fluke 8840A Language
GPIB / 488
X
X
X
RS-232
X
Not allowed
Not allowed
Chapter 3 Features and Functions
Remote Interface Configuration
92