32 KODAK Display Manager System V4.1 User’s Guide
Calibration
The Calibration Procedure
To automatically calibrate a monitor:
1. Click Auto from the drop-down Display Mode menu if it isn’t already
selected.
NOTE
Auto mode is only available if you have connected the enable cable to your
computer.
2. Click the Calibrate icon.
If this is the first time that you have calibrated your current monitor, follow
the on-screen prompts to run a monitor evaluation. See
Monitor
Evaluation” on page 35.
Either the Calibration window opens or you are prompted to perform a
Dark Gain Correction to reset the sensor.
If you are prompted to perform a Dark Gain Correction:
a) Place the sensor on a dark surface and make sure that no light can
reach it. Do not attach the sensor to the monitor for this correction.
b) Click OK.
IMPORTANT
If the Dark Gain Correction fails, see Dark Gain Correction Failure” on
page 78.