Princeton LS 785 Camera Lens User Manual


 
22 LS 785 High Throughput Lens Spectrograph User Manual
PIXIS
Tools:
5/32hex wrench
5/16” wrench
Procedure:
1. If detector other than a PIXIS was previously mounted to the LS 785, re-install the four #8-32 socket head
screws and washers used in the instrument housing to secure the other adapter.
2. If the spacer is not installed, mount it to the face of the LS 785 with three (3) #10-32 socket head screws
(5/32” hex wrench).
3. Attach the PIXIS adapter to the LS 785 with label at the top with three #10-32 socket head screws (5/32”
hex wrench).
4. In three locations (indicated), screw in three 3/8”x#10-32 screws provided with the detector. Leave
approximately 3/16” thread exposed.
5. If an O-ring is supplied with the PIXIS, insert it into the groove in the outside face of the PIXIS flange.
6. Mount the PIXIS detector to the adapter and tighten bolts.
Label
“Top”
3 xs #10-32
Socket Head
Screws
PIXIS CCD
Adapter
3 x #10-32
Flat Head Bolts