Omega Vehicle Security DP63000A-I Camera Accessories User Manual


 
Entering Code 66 will overwrite all user settings with
the factory settings. The meter will display rESEt and then
return to CodE 00. Press the SEL button to exit the module.
10
6.2 MODULE 2 - SECONDARY FUNCTION PARAMETERS (2-SEC)
PARAMETER MENU
MIN DISPLAY ENABLE
USER INPUT ASSIGNMENT
Select the value(s) to which the User Input Function is assigned. The User
Input Assignment only applies if a selection of reset, display hold, or print and
reset is selected in the User Input Function menu.
MODEDISPLAY
No Function
NO
DESCRIPTION
User Input disabled.
Program Mode Lock-out
P-Loc
Zero Input
(Edge triggered)
ZErO
Zero the Input Display value causing
Display Reading to be Offset.
COLOr
d-SEL
d-LEV
d-HLd
rESEt
Resets the assigned value(s) to the
current input value.
HI-LO HI
dSPLO
See Programming Mode Access chart
(Module 3).
YESNO
MODE DESCRIPTIONDISPLAY
Setpoint 1 and 2 Reset
Setpoint 1 Reset
Setpoint 2 Reset
Print and Reset
Reset both setpoint 1 and 2 outputs.
Resets setpoint 1 output.
Resets setpoint 2 output.
Same as Print Request followed by a
momentary reset of the assigned value(s).
rSt12
rSt-1
rSt-2
P-r5t
Print Request
Serial transmit of the active parameters
selected in the Print Options menu
(Module 5).
Print
ª
«
2.0
HI-t
ª
«
NO
LO-En
ª
«
NO
FCS
MAX CAPTURE DELAY TIME
When the Input Display is above the present MAX value for the entered
delay time, the meter will capture that display value as the new MAX reading.
A delay time helps to avoid false captures of sudden short spikes.
ª
«
2.0
LO-t
MIN CAPTURE DELAY TIME
When the Input Display is below the present MIN value for the entered delay
time, the meter will capture that display value as the new MIN reading. Adelay
time helps to avoid false captures of sudden short spikes.
0.0 to 999.9 sec.
USER INPUT FUNCTION
ª
«
NO
USrIN
ª
«
dSP
U-ASN
ª
«
NO
HI-En
MAX DISPLAY ENABLE
0.0 to 999.9 sec.
Select yES to perform either of the Factory Service Operations shown below.
FACTORY SERVICE OPERATIONS
yESNO
YESNO
Enables the Maximum Display Capture capability.
Enables the Minimum Display Capture capability.
The unit uses stored current calibration values to
provide accurate current measurements. Over time, the
electrical characteristics of the components inside the unit
will slowly change with the result that the stored
calibration values no longer accurately define the input circuit. For most
applications, recalibration every 1 to 2 years should be sufficient.
Calibration of the unit involves a current calibration which should only be
performed by individuals experienced in calibrating electronic equipment.
Allow 30 minute warm up before performing any calibration related procedure.
The following procedures should be performed at an ambient temperature of 15
to 35 °C (59 to 95 °F).
CAUTION: The accuracy of the calibration equipment will directly affect the
accuracy of the unit.
Current Calibration
1. Connect the negative lead of a precision DC current source with an accuracy
of 0.01% or better to the COMM terminal. Leave the positive lead of the DC
current source unconnected.
2. With the display at CodE 48, press and hold the SELbutton for 2 seconds. Unit
will display CAL NO
3. Press the RST button to select the range to be calibrated.
4. Press the SEL button. Display reads 0.0A
5. With the positive lead of the DC current source unconnected, press SEL.
Display reads CALC for about 8 seconds.
6. When the display reads the selected range, connect the positive lead of the DC
current source to INP+ and apply full-scale input signal for the range. (Note:
For 200 mA range, apply 100 mAas indicated on the display.)
7. Repeat steps 3 through 6 for each input range to be calibrated. When display
reads CAL NO, press the SEL button to exit calibration.
CALIBRATION
ª
«
48
CodE
ª
«
66
CodE
RESTORE FACTORY DEFAULT SETTINGS
Backlight Color
(Edge Triggered)
Display Select
(Edge Triggered)
Display Intensity Level
(Edge Triggered)
Display Hold
Reset (Edge triggered)
Change backlight color with each
activation (backlight version only).
Advance once for each activation.
Increase intensity one level for each
activation (backlight version only).
Holds the assigned display, but all other
meter functions continue as long as
activated (maintained action).
Entering Code 50 will display the version (x.x) of the
meter. The display the returns to CodE 00. Press the SEL
button to exit the module.
ª
«
50
CodE
VIEW VERSION DISPLAY