89
FUNCTIONAL DESCRIPTION HI931000
ins rum n
s
CFM
°C
CAL
OFF
ON
HI 931000
1
2
3
4
5
6
1) DIN socket for combination pH/°C electrode.
2) Liquid Crystal Display (LCD).
3) ON/OFF key, to switch the instrument on or off.
4) °C key, to display the measured temperature (press & hold).
5) CAL key, to enter calibration mode.
6) CFM key, to confirm calibration data.
SPECIFICATIONS HI931000
HI 931000
Range pH 0.00 to 14.00
°C 0.0 to 70.0
Resolution pH 0.01
°C 0.1
Accuracy pH ±0.01
(@20°C/68°F) °C ±0.5
Typical EMC pH ±0.30
Deviation °C ±1
pH Calibration Automatic 2 point with
3 memorized standard buffers
(pH 7.01, 4.01, 10.01)
Temperature Automatic from 0 to 70°C
Compensation (32 to 158°F)
pH Electrode HI 1217D combination, gel
with built-in temperature sensor
and amplifier (included)
Battery Type 1 x 9V
Life 300 hours of continuous use
Auto-off after 8 minutes of non-use
Environment 0 to 50°C (32 to 122°F);
max 95% RH non-condensing
Dimensions 143 x 80 x 38 mm
(5.6 x 3.2 x 1.5")
Weight 360 g (12.7 oz.)