Princeton NFC-446-040 Camera Accessories User Manual


 
Figure 1: NFC-446-040 Layout (Top View)
Mounting the NFC-446-040 to Acton Series Monochromators or Spectrographs:
Use the illustration above for reference when mounting the NFC-446-040 to ARC
monochromators or spectrographs.
1. Unpack the NFC-446-040 Raman Notch Filter Chamber.
2. Remove the cover of the NFC-446-040.
3. Set the components on a stable, flat working surface. Position the NFC-446-040
against the ENTRANCE SLIT of the Acton Series monochromator or spectrograph
as shown in the illustration above. Align the mounting holes of the NFC-446-040
with the tapped holes in the slit assembly.
4. Insert the two (2) 8-32 screws provided through the mounting holes and into the
tapped holes of the slit assembly. Tighten to secure the NFC-446-040 to the
entrance slit.
5. The NFC-446-040 includes a support post to help stabilize the system. This post
should rest flat on the working surface, and should not exert stress on the system.
The NFC-446-040 is now mounted to the Acton Series monochromator or
spectrograph.