54
FUNCTIONAL DESCRIPTION &
SPECIFICATIONS OF HI 9033
1) Battery compartment
2) Probe socket
3) Liquid Crystal Display
4) Measure unit
5) Low battery indicator
6) ON/OFF key
7) 19.99 mS/cm range
selection key
8) Calibration trimmer
9) 199.9 μS/cm range
selection key
10) 1999 μS/cm range
selection key
11) 199.9 mS/cm range
selection key
FUNCTIONAL DESCRIPTION &
SPECIFICATIONS OF HI 9034
1) Battery compartment
2) Probe socket
3) Liquid Crystal Display
4) Low battery indicator
5) ON/OFF key
6) 19.99 g/L range
selection key
7) Calibration trimmer
8) 1999 mg/L range
selection key
9) 199.9 mg/L range
selection key
Range 0.0 to 199.9 mg/L / 0 to 1999 mg/L
0.00 to 19.99 g/L
Resolution 0.1 mg/L / 1 mg/L / 0.01 g/L
Accuracy (@20°C/68°F)
±1% F.S. (excluding probe error)
Typical EMC Deviation ±1% F.S.
Calibration Manual, 1 point, through trimmer
Temperature Compensation
Automatic, 10 to 50°C (50 to 122°F) with ß=2%/°C
Probe HI 76302W with 1 m (3.3') cable (included)
Battery Type / Life
1 x 9V / APPROX. 100 hours of continuous use
Environment 0 to 50°C (32 to 122°F); RH max 100%
Dimensions 196 x 80 x 60 mm (7.7 x 3.1 x 2.4")
Weight 425 g (0.9 lb.)
Range 0.0 to 199.9 μS/cm / 0 to 1999 μS/cm
0.00 to 19.99 mS/cm / 0.0 to 199.9 mS/cm
Resolution 0.1 μS/cm / 1 μS/cm
0.01 mS/cm / 0.1 mS/cm
Accuracy (@20°C/68°F)
±1% F.S. (excluding probe error)
Typical EMC Deviation ±1% F.S.
Calibration Manual, 1 point, through trimmer
Temperature Compensation
Automatic, 10 to 50°C (50 to 122°F) with ß=2%/°C
Probe HI 76302W with 1 m (3.3') cable (included)
Battery Type / Life
1 x 9V / APPROX. 100 hours of continuous use
Environment 0 to 50°C (32 to 122°F); RH max 100%
Dimensions 196 x 80 x 60 mm (7.7 x 3.1 x 2.4")
Weight 425 g (0.9 lb.)