A SERVICE OF

logo

1110
Tap the probe repeatedly on the bottom of the beaker and stir it
to ensure that no air bubbles remain trapped inside the sleeve.
If the probe temperature is close to the solution temperature, the
display will stabilize quickly and show the temperature compen-
sated conductivity measurement.
However, the ATC circuitry needs a few minutes for temperature
compensation if the temperature difference is about 5°C (9°F).
When the reading is stable, adjust the calibration trimmer to
read the solution conductivity value @25°C.
For example if using HI 7032, turn the trimmer until the display
shows 1382 mg/L @25°C.
The calibration is now complete and the instrument is ready for use.
All subsequent measurements will be compensated to 25°C (77°F).
Note: Calibration can also be standardized to have readings com-
pensated to 20°C (68°F), by simply adjusting the trimmer to
the calibration solution value at 20°C (68°F). For example, for
HI 7032 adjust the trimmer to read 1251 mg/L (see "TDS vs.
Temperature" section).
If the instrument cannot be calibrated, refer to the "Probe Mainte-
nance" section.
When the reading is stable, adjust the calibration trimmer to
read the solution conductivity value @25°C.
For example if using HI 7030, turn the trimmer to read
12.88 mS/cm @ 25°C.
The calibration is now complete and the instrument is ready for use.
All subsequent measurements will be compensated to 25°C (77°F).
Note: Calibration can also be standardized to have readings com-
pensated to 20°C (68°F), by simply adjusting the trimmer to
the calibration solution value at 20°C (68°F). For example, for
HI 7030 adjust the trimmer to read 11.67 mS/cm (see
"Conductivity vs. Temperature" section).
If the instrument cannot be calibrated, refer to the "Probe Mainte-
nance" section.
PROCEDURE FOR HI 9034
Fill a beaker with 8 cm (3¼") of TDS
calibration solution (if possible fill two
beakers, one for rinsing the probe and
one for calibration).
Connect the probe to the meter securely
by aligning the pins with the socket,
pushing the plug in and tightening the
threaded ring.
Immerse the probe into the beaker, while
paying attention that the holes on the
PVC sleeve are completely submerged.
Turn the instrument on by pressing the ON/OFF key and select
the appropriate range (e.g. 1999 mg/L).
ON/OFF
1999mg
/l
mS