Panasonic WV-CLR924AE Digital Camera User Manual


 
-39-
6. Set the oscilloscope to the horizontal rate and
expand the horizontal sync portion on the oscillo-
scope.
7. Move the cursor to H PHASE.
8. Adjust the horizontal phase by pressing
L or M.
The adjustable range is 0 - 2.0 µs.
6-4. Line-lock Sync Mode (LL)
Note:
The line-lock mode is not applicable when
DC power and/or VD2 synchronization is sup-
plied to the camera.
1. Move the cursor to SYNC and select LL.
2. After confirming that the cursor is on LL, press
I.
The vertical phase adjustment menu appears on
the monitor.
3. Supply the video output signal of the camera to be
adjusted and the reference camera video output
signal to a dual-trace oscilloscope.
4. Set the oscilloscope to the vertical rate and expand
the vertical sync portion on the oscilloscope.
5. Move the cursor to COARSE. The cursor is high-
lighted.
6. Press
L or M to match the vertical phase for both
video output signals as closely as possible.
(COARSE adjustment can be incremented in 16
steps by 22.5 degrees by pressing
L or M.)
Note: After the sixteenth step, the adjustment
returns to the first step.
7. Move the cursor to FINE.
8. Press
L or M to match the vertical phase for
both video output signals as closely as possible.
(FINE adjustment can be made up to 22.5 degrees
by pressing
L or M.)
1 (1 - - - 16): 0 degrees
2 (1 - - - 16): 22.5 degrees
16 (1 - - - 16): 337.5 degrees
** SYNC **
V PHASE
COARSE 1(1--16)
FINE I........
- +
RET END