Making Basic Measurements Using Spectrum Analyzer Mode
Tracking Unstable Signals Using Marker Track and the Maximum Hold and Minimum Hold Functions
6.
Press (TRACE_), TRACE A 3 C to select trace B. (Trace B is selected when
B is underlined.) Press CLEAR WRITE
B
to place trace B in clear-write
mode, which displays the current measurement results as it sweeps.
Trace A remains in maximum-hold mode, showing the frequency shift of
the signal.
7. Press TRACE A 3 C to select trace C (C should be underlined). Press
MIN HOLD C Trace C is in the minimum-hold mode and displays the
minimum amplitude of the frequency drift of the signal.
&4.
Ii.
46 JUL
29,
1993
MKA-TRK
100
"963
MHZ
REF
9.0
dBm”
#AT
0
dB
_
7.93
dmn”
PEAK
LOG
10
dE/
Figure
4.10.
Viewing an Unstable Signal With Max Hold, Clear Write, and Min Hold
4-17