Step 1
Turn on the power. The indicator light on the power switch will come on.
The pump starts to pump air into the mattress.
TheLowPressureindicator(yelloworred)willcomeonasinationundergoes.
Theaudiblealarmisactivated,accompanyingwiththeLowPressureindicator,toalertthatpressureis
low.PresstheAlarmResetbuttontomutetheaudiblealarm,anditsindicatorwillash.
NOTE! For first inflation, we recommend that you press both the Alarm Reset and Static buttons to enter
a quick and silent operation.
Step 2
Whenthepressurereachesthepresetlevelbyfactorydefaultwithinapproximately30~45minutes,the
NormalPressureindicatorwillcomeon,andtheLowPressureindicator(andtheaudible/visiblealarm)
will turn off.
NOTE! If the Low Pressure indicator (and the audible/visible alarm) won’t go off, please refer to page 15
for troubleshooting.
Step 3
Turnthe pressure-adjustknobtosetacomfortablepressurelevelfromsofttorm.(Pressurerange:
30~60mmHgforQ2-02).
NOTE! For suitable pressure, please refer to page 9 for the hand check procedure.
Step 4
The pump is preset in alternating mode and its cycle time is set at 10 min/60Hz or 12 min/50Hz.
Press the Static button to set it in static mode, and the Static indicator will come on. The static mode will
bestartedwithinapproximately6minutes.PresstheStaticbuttonagaintoswitchbacktoalternating
mode.
NOTE! In static mode, the mattress provides a firm surface that makes it easier for the patient to transfer
or reposition. The static mode prevents the patient from bottoming out when in a sitting position.
Step 5
Duringnormal operation,the pump will monitorpressure changes. If the pressure isbelow a user-
denedpressurelevel,thepumpwillautomaticallystarttoinatethemattress.Accompanyingwiththe
Low Pressure indicator, the audible alarm will beep and its indicator will come on to bring attention to a
lowpressurecondition.Thepumpwillstopwhentheuser-denedpressurelevelisreached.
PresstheAlarmResetbuttontomutethealarm;itsindicatorwillash.
PresstheAlarmResetbuttonagaintohavetheaudiblealarmanditsindicatoron.