JVC DZ-VCA1SE Security Camera User Manual


 
11
Connections
Connection example
Notes:
Do not connect this camera to equipment other than that depicted in these connection diagrams.
Always be certain to use the AA-V31E Isolation Transformer.
The camera head and camera control unit are adjusted as a set. Be sure the serial numbers match.
Optimum picture quality may not be obtained if a different combination is used.
When the system is connected, check that current leakage from this connection does not surpass the specified
level. Also, be absolutely sure to use an AA-V31E isolation transformer.
If the exterior of the isolation transformer is corroded, inadequate grounding may result. Have the exterior
replaced by your local JVC dealer.
RGB signal HD video signal
(Y, PB, PR)
Camera control unit (rear panel)
Front panel
DC 12 V
AC power
adapter
(AA-V112E)
AC 230 V
Camera head
Lens
(commercially
available)
1/2" bayonet mount
conversion adapter
(GL-V62J)
Camera head
Microscope
adapter GL-V61J
(optional)
Camera cable
VC-V10 or VC-V3
(optional)
Isolation transformer
AA-V31E
HD Monitor
HV-M2000V
W-VHS VCR SR-W7 MAE
Endoscope
Endoscope Video Adapter GL-V63J*
Endoscope Video Zoom Adapter GL-V64J*
* Always use an adapter that is designed for the Endoscope.
DC-IN
GEN-LOCK IN
SYNC OUT
V S/ C. SYNC
(
75
)
HD
VD
C. SYNC
HD OUT
G
B
Y
P
B
RPR
CAMERA
ZIP drive
Z250S
Endoscope application Operation field application
Operation microscope application