Sigma LBA-700 Digital Camera User Manual


 
Centroid X,
Centroid Y,
Width X,
Width Y,
Diameter(^END)
LBA-PC sends
RDR
,%,,,um,um,um,um,um,um,um(^END)
10.6.2.6 LOG - Logging
You can configure the LBA-PC to automatically send a FRM or RDD and RDR response each time
new data is acquired. Remote logging is the fastest method available to transfer new frames
from the LBA-PC.
Use the File | Logging… dialog box or the LOG configuration command to specify logging new
data or results to the host. If the data log file name is “FRM” or “RDD” then each time a new
frame of data is acquired the LBA-PC will automatically send a response as if you had sent
“:FRM?” or “:RDD?” respectively. If the results log file name is “RDR” then each time a new
frame of data is acquired the LBA-PC will automatically send a response as if you had sent
“:RDR?”.
The LBA-PC will not capture another frame until the host reads the current logging response.
This example describes how to enable data file logging to the host controller.
Host sends
:LOG
DataLogging=1;
DataFileName=FRM(^END)
Host sends
:RUN
LBA-PC sends
FRM
FrameNumber=1;
#516384(DAB)…(DAB) (^END)
LBA-PC sends
Operator’s Manual LBA-PC
Doc. No. 10654-001, Rev 4.10
197
FRM
FrameNumber=2;