24 Advanced Camera Operation Manual
PVCAM Class 101: ICL pl_exp_setup_script(101)
NAME
pl_exp_setup_script – process a script; download instructions to the
camera.
SYNOPSIS
boolean
pl_exp_setup_script (int16 hcam, char_const_ptr script,
uns32_ptr stream_size, uns32_ptr num_rects)
DESCRIPTION
hcam
must specify a CCD camera that has been successfully opened by PVCAM.
This function uses the input string
script
to create a control program for
hcam
.
The script controls all aspects of one or more exposures, including clearing the
CCD, opening and closing the shutter, shifting the parallel register (or registers,
in the case of frame transfer CCDs) and reading out the pixels. The script is also
capable of coordinating with external triggers and triggering external hardware.
This function compiles the script into a form that is appropriate for control-ling
hcam
. During that compilation, it determines how many pixels will be collected.
The total number of bytes is returned in
stream_size
; and the user must
allocate a
pixel_stream
array of at least this size before running the script (see
pl_exp_start_script).
During compilation, the scripts display functions are also examined. These
functions specify how the resulting data is decomposed into individual images.
After this function has successfully returned, the caller is assured that the script
is accepted and properly processed, the corresponding instructions have been
transmitted to the camera (which is now ready to go), and the display
instructions have been processed. Data collection may now start by calling
pl_exp_start_script function.
RETURN VALUE
TRUE for success, FALSE for a failure. Failure sets pl_error_code.
SEE ALSO
pl_exp_start_script(101), pl_exp_abort(0),
pl_exp_check_status(0)
NOTES
Most of the complexity of this function is contained in the scripting
language. Read this manual for a more complete description of this
function.