A SERVICE OF

logo

Configuration and Use Manual 215
Diagnostics and Troubleshooting
Diagnostics and Troubleshooting Specifications – Model 3350/3700Specifications – Model 3300/3500Measurement Performance
22.7.2 Sensor alarms
Table 22-5 lists all alarms in the Sensor category, with descriptions, suggested user actions, and
related information.
A130 Printer Out of
Paper
(Custody transfer
only) The ticket
printer is out of
paper.
Ignore Yes No Add paper to the printer.
A134 PPI Memory
Error
Memory or
checksum error
in transmitter
display.
Info Yes No Cycle power to the
flowmeter. If the problem
persists, contact Micro
Motion customer service.
A135 PPI Comm Error Internal
communication
error with
transmitter
display.
Info Yes No Cycle power to the
flowmeter. If the problem
persists, contact Micro
Motion customer service.
(1) Applies only to systems with the standard core processor.
(2) Applies only to systems with the enhanced core processor.
(3) Applies only to systems with transmitter software earlier than rev7.0.
Table 22-5 Sensor alarms
Alarm
number
Maintenance
menu listing Description
Alarm severity
Affected
by fault
timeout User actionsDefault
User
config?
A003 Sensor Failure No tube vibration
detected at the
sensor.
Fault Yes Yes Check the test points. See
Section 22.18.
Check the sensor coils. See
Section 22.20.
Check wiring to sensor. See
Section 22.11.2.
Check for slug flow. See
Section 22.7.3.
Check sensor tubes.
Table 22-4 Electronics alarms continued
Alarm
number
Maintenance
menu listing Description
Alarm severity
Affected
by fault
timeout User actionsDefault
User
config?