Series 3000 MVD Transmitters and Controllers
Diagnostics and Troubleshooting
22.7.3 Process alarms
Table 22-6 lists all alarms in the Process category, with descriptions, suggested user actions, and
related information. More information on specific process alarms is provided later in this section.
A004 Temp.
Out-of-Range
The measured
temperature is
outside the
sensor limits.
Fault No Yes Check wiring to sensor. See
Section 22.11.2.
Verify sensor configuration.
See Section 7.3.3.
Check the test points. See
Section 22.18.
Check the sensor coils and
RTD. See Section 22.20.
Verify that process
temperature is within range
of sensor and transmitter.
Contact Micro Motion
customer service.
A016 Sensor RTD Error Sensor RTD has
failed.
Fault Yes Yes Check wiring to sensor. See
Section 22.11.2.
Make sure the appropriate
sensor type is configured.
See Section 7.3.3.
Check the test points. See
Section 22.18.
Check the sensor coils. See
Section 22.20.
Contact Micro Motion
customer service.
A017 Meter RTD Error Meter RTD has
failed.
Fault Yes Yes Check wiring to sensor. See
Section 22.11.2.
Make sure the appropriate
sensor type is configured.
See Section 7.3.3.
Contact Micro Motion
customer service.
Table 22-5 Sensor alarms continued
Alarm
number
Maintenance
menu listing Description
Alarm severity
Affected
by fault
timeout User actionsDefault
User
config?