Fujitsu fi-5120C Photo Scanner User Manual


 
fi-5120C / fi-5220C Image Scanner Operator
Guide
TROUBLE SHOOTING
CONTENTS MAP
TOP
4.3 Cleaning the ADF (4)
1 2 3
4
Feed roller
Clean lightly taking care not to roughen the surface of the rollers. Take particular care in cleaning the feed rollers since
black debris on the rollers adversely affect the pickup performance.
1. Open the ADF when the indication of the Function Number display is not "P" or "0".
If you open the ADF when the indication of the Function Number display is "P" or "0", the Feed rollers will not rotate
even you perform the operation of step 2 below.
2. Simultaneously hold down the "Send to" and "Scan" buttons on the operator panel.
The Feed rollers start to rotate slowly.
3. Hold a soft, cloth moistened with cleaning fluid against the surface of the rotating Feed rollers so that it lightly
cleans the surface of the rollers. As a guideline, seven presses of the "Send to" and "Scan" buttons make the Feed
rollers to rotate one full turn.
:Button "Send to"
:Button "Scan"
CAUTION
In step 2, all the Feed rollers turn at the same time. When cleaning the Feed rollers
positioned in the lower part of the ADF, be caereful not to touch the other Feed rollers.
With the movie below, review the procedure to clean the Feed rollers:
3. Press down on the center of the ADF to return it to its original position until the ADF button locks.