Princeton LS 785 Camera Lens User Manual


 
24 LS 785 High Throughput Lens Spectrograph User Manual
Spec 10 LN without Shutter
Tools:
9/64” hex wrench for removing 4 screws with washers.
5/32” hex wrench for removing three socket heads holding spacer on.
Procedure:
1. If another adapter is mounted to the LS 785, remove that adapter (and, in the case of the PIXIS, the
spacer).
2. Using a 9/64” hex wrench, remove the four #8/32 socket head screws with washers from the face of the
LS 785.
3. Locate the Spec 10 LN detector without a shutter and mark the top of its mounting flange.
4. Remove the mounting flange from the Spec 10 LN without a shutter.
5. Align the tops of the Spec 10 mounting flange and the adapter flange. Attach the Spec 10 mounting flange
to the adapter with six #10-32x 1/8” flat head screws.
6. Re-attach the Spec 10 mounting flange and adapter assembly to the Spec 10 by installing the four screws
through the four clearance holes in the adapter.
7. Attach the Spec 10 and adapter to the LS 785 with the four #8-32 socket head screws and washers where
shown in the photographs.
4 x #8-32
Socket Head
Screws with
Washers
Label
“Top”
4 x #8-32
Socket Head
Screws with
Washers