Year 2000 Conversion Instructions and Test Procedure
OLYMPUS REPLYand OLYMPUS AU560 Automated Chemistry Analyzers
Year 2000 Conversion Instructions:
Note: If you wish to do a year 2000 test of the instrument please see page 5.
The following items are required to perform this procedure on the Reply/AU560.
Program Floppy Disk 1/3
MU5907 SPECIAL TOOL for YEAR 2000 diskette
1. Confirm that POWER ON switch on the front of the DPR is not illuminated and the analyzer
and DPR are off.
2. Move the BOOT switch on the rear side of the DPR from “WD” to “FD”. When moving the
switch, please pull out on the switch. If the switch is not pulled, it cannot be moved (FIG.12-
¬-).
3. Switch ON the LOCAL POWER switch on the rear side of the DPR (FIG.12-¯). Wait until
the DPR displays the message ”Please insert system floppy disk & hit any key then restart disk
mount”.
4. Confirm that Write Protect is off on the Program Floppy Disk 1/3, insert it into the FD drive
and press the Enter key.
5. Wait until the DPR displays the following menu (FIG.1).
*** SYSTEM MAINTENANCE ***
# SELECT No.>
---------------------------------------------------------------------
1 INITIALIZE MEDIA
2 PASSWORD SET
3 COMMUNICATION
4 MAINTENANCE
5 HISTORY
6 COPY FILES
7 SYSTEM UP
1: 2: 3: 4: 5: 6: 7: 8:EXIT
FIG.1
6. Type 7 and the Enter key on the keyboard.
7. Wait until the DPR displays the following menu (FIG.2). Only Item 1 displays at this time.