Canon MLT (F) Camera Accessories User Manual


 
2.Edit Exposure Mode Settings
– 8 –
3
Display the FREQUENCY BASE window
Touch the FREQUENCY BASE button to open the window.
FREQUENCY BASE Window
The types of Frequency Bases are classified into two kinds, and are indicated with the
buttons in different colors respectively as follows:
The bar graph in the each button shows the range of frequencies that enhancement
applies to.
To enhance the mass structure, such as soft tissues, select the Frequency Base that
applies enhancement to the lower frequency. The range increases by including the lower
frequency in ascending order from A or a. To enhance the minute structure, such as bone
tissues, select the Frequency Base that limits enhancement to the lower frequency and
covers only the higher frequency. The range decreases by excluding the lower frequency
in descending order to A or a.
4
Set the FREQUENCY BASE parameter
Touch the desired Frequency Base button.
The parameter is set for the Frequency Base selected, and the FREQUENCY BASE win-
dow is closed.
Display of Button Remarks
Blue Bar Graph
Enhances the signals smoothly ranging from the low frequency
area to the high frequency area.
The results of enhancement appear natural with smooth frequent
response characteristics.
Green Bar Graph
Enhances the structures in the specific frequency band stronger
than the structure in other frequency bands.