Canon MLT (F) Camera Accessories User Manual


 
– 29 –
4.QA Parameters
9 Suppress Grid Processing
One of the following parameters is displayed:
Z0 (No grid stripe is detected), Z1 (Grid stripes are suppressed), Z* (Processing is
not determined yet)
If the Grid Reduction Processing is not effective, nothing is displayed.
10 High-Density Clipping Processing (HIGH IN. CLIPPING)
One of the following parameters is displayed:
C0 (Clipping is not performed), C1 (Clipping is performed), C* (Processing is not
determined yet)
If the High-Density Clipping Processing is not effective, nothing is displayed.
11 Dose-Depend Processing (D. DEPEND MTF IMP)
One of the following parameters is displayed:
M0 (Processing is not performed), M1 (Processing is performed in the fix mode),
M2,n,n (Processing is performed in the dose dependent mode)
The parameters for M2 are displayed as follows:
M2, n (FILTER TYPE Index), n (INTENSITY Index)
The display values correspond to the parts of the body as shown in the following
table (the values for INTENSITY are omitted):
FILTER TYPE
Display Value1234567
Body Part Chest
Abdo-
men
Skull
Cervical
vertebrae
Thoracic
vertebra
Lumbar Limbs