Maintaining the Equipment
August 2001 4-3
Cleaning the Light Bar
If artifacts appear on your scanned images, clean the light bar.
IMPORTANT: Wear white cotton gloves for this procedure to prevent oils and dirt
from being absorbed into the microfiber cleaning cloth.
1. Fold a dry, untreated microfiber cleaning cloth so the edge of the cloth is
about the thickness of the light bar.
NOTE: You can use an e-wipe, manufactured by Photographic Solutions, Inc.,
instead of a microfiber cleaning cloth.
2. Gently move the edge of the microfiber cloth or e-wipe down the length of the
light bar one time.
3. Calibrate the scanner.
4. If an artifact remains after cleaning the light bar, check the light bar with a
magnifying glass for damage, such as a scratch or a chip. If there is damage
to the light bar, contact your Kodak representative for service.