3.2.11 Password Lockout
You can enter a Password that will cause all of the LBA-PC setup functions to become inaccessible.
The password acts as a toggle. Type in the password once and all setup options will become
disabled. Type it in a second time and setup capability will be restored.
The password has been factory set. To find out your password call the Spiricon Service Department
at 435-753-3729 between the hours of 9:00 am and 5:00 pm MST. The password will only be given
out to an authorized individual. An authorized individual is one whose name appears on the original
Purchase Order, the person whose name appears on the completed warranty card, or to an individual
who knows the Purchase Order Number used to make the purchase. It can also be obtained by a
written request on your company letterhead, signed by an officer of the company.
3.3 Pass/Fail... Drop Down Menu Selections
3.3.1 Pass/Fail... enable and define its operation
All pass fail operations are enabled or disabled by checking the Master Enable box.
Figure 42
You can choose what Action(s) will be performed based upon if the results Fails or Passes the
designated tests.
Check the appropriate box to obtain these available actions:
TTL Pulse: If you want the LBA to send a TTL pulse out the rear J5 connector, pin 1, each
time the Pass/Fail condition is met.
Beep: If you want the LBA to cause your PC to Beep, each time the Pass/Fail condition is
met.
Stop Running: If you want the LBA to halt and display the frame that caused the Pass/Fail
condition to be met.
Operator’s Manual LBA-PC
98