SCALE OPERATION
CALIBRATION
Bar Code Calibration Procedure without Remote Display*
8. Place a FULL Load on the center of the scale platter. The white LEDs will blink 3 times periodically. Wait at least 3 seconds for
scale stability, and then press the Multi-Function Button once. After the Multi-Function Button has been pressed, there will be a
short delay followed by three short beeps indicating that the FULL Load value has been stored.
Figure 52.
9. Remove the entire load from the scale platter. The white LEDs will blink 5 times periodically. Wait at least 3 seconds for scale
stability, and then press the Multi-Function Button once. After the Multi-Function Button has been pressed, there will be a short
delay followed by five short beeps indicating that the zero load value has been stored.
Figure 53.
10. Calibration process is now complete. The scanner will automatically restart, and beep one time, in about 12 seconds.
11. Verify the calibration was successful with the calibrated weight. If the scale display does not accurately show the calibrated load
weight correctly, restart this procedure from Step 1 on page 42.
12. The unit’s calibration must now be verified as required by state and/or local Weight and Measures regulations
(starting on Page 46).
* These procedures are for MS2320 StratosH Scanner/Diva Scale models only.