Leica SP2600 Digital Camera User Manual


 
47
Leica SP 2600  Ultramilling attachment
7.2.2 Replacing the light bulb in the calibration device
Remove base cover (2) of calibration device (1) (unscrew 4 slotted
screws (3)).
Loosen light bulb holder (5) (metal bridge with lamp socket) (unscrew 2
slotted screws (4)).
Take replacement light bulb (8) out of sponge rubber cube (7).
Remove broken light bulb (6) from lamp socket and replace with new
light bulb (8).
Place another replacement bulb (8) into the sponge rubber cube (7).
Put light bulb holder (5) back in place.
Place ring with cable (9) onto bore (10) in the light bulb holder.
Fasten light bulb holder with two screws (4).
Put sponge rubber cube (7) with new replacement light bulb (8) back into
the calibration device.
Put base cover (2) back in place and fasten with 4 screws (3).
7. Cleaning, disinfection and maintenance
Fig. 7.1
7
8
2
3
4
6
5
10
9
1