Konica Minolta CS-200 Camera Accessories User Manual


 
108
Description
Operation Procedure (With data management software CS-S10w)
Setting Necessary for Object Color Measurement
1.
Set one or more tungsten lumps
or equivalent as illumination
source toward white calibration
plate as in the right illustration.
Set this instrument vertical
to white calibration plate.
Keep the angle between
illumination light source and
white calibration plate at 45˚.
2.
Start up this instrument, PC, and software CS-S10w.
White Calibration
3.
Set color measuring mode for object color with CS-S10w.
4.
Conduct white calibration with CS-S10w.
Measurement of Object
5.
Set object on the same position and at the
same angle as white calibration plate.
6.
Now, measurement can be done with CS-S10w.
See instruction manual of data management software CS-S10w for details.
Illumination
Light Source
Illumination
Light Source
White calibration plate