Error message Cause (Description) Corrective Action
7
OVER
Luminance of measuring
object is higher than avail-
able measurement range.
• Use ND filter and redo measure-
ment.
• If symptom does not improve,
please contact the nearest KONI-
CA MINOLTA SENSING authorized
service facility.
8
OFFSET ERROR
Zero calibration has not
been correctly performed.
• Reset the power and redo meas-
urement.
• If symptom does not improve,
please contact the nearest KONI-
CA MINOLTA SENSING authorized
service facility.
9
EXCESSIVE LUMI-
NANCE VARIATION
Luminance change for
measuring object is large.
• Place this instrument on stable
surface and redo measurement.
• If symptom does not improve,
please contact the nearest KONI-
CA MINOLTA SENSING authorized
service facility.
10
VIEWING-ANGLE
SELECTOR ERROR
Measurement was per-
formed when measurement
angle selector was in wrong
position, or measurement
angle selector was changed
during measurement.
• Switch measurement angle selec-
tor and check that measurement
angle appears in screen (------ is
not displayed), and then redo
measurement. In addition, do not
operate measurement angle selec-
tor during measurement.
• If symptom does not improve,
please contact the nearest KONI-
CA MINOLTA SENSING authorized
service facility.
11
ROM ERROR
Data stored in ROM is
broken.
• Do not switch OFF (O side) while
storing data or changing setting, or
when message “PLEASE WAIT...”
appears.
• If symptom does not improve,
please contact the nearest KONI-
CA MINOLTA SENSING authorized
service facility.