Konica Minolta CS-200 Camera Accessories User Manual


 
112
Description
Error message Cause (Description) Corrective Action
12
MEMORY
ERROR
Data stored in ROM is
broken, or backup batteries
drain.
• Do not switch OFF (O side) while
storing data or changing setting, or
when message “PLEASE WAIT…”
appears.
• Switch ON (| side) to charge back-
up battery. Backup battery is to
fulfill its function after fully charged
for approx. 20 hours.
• If symptom does not improve,
please contact the nearest KONI-
CA MINOLTA SENSING authorized
service facility.
13
INCORRECT
CLOCK
OPERATION
Clock IC does not operate
correctly.
• Reset the power to set correct
date and time.
• If symptom does not improve,
please contact the nearest KONI-
CA MINOLTA SENSING authorized
service facility.
14
A/D ERROR Error in AD converter • Reset the power to redo measure-
ment.
• If symptom does not improve,
please contact the nearest KONI-
CA MINOLTA SENSING authorized
service facility.
15
RECALCULA-
TION ERROR
There exists no measure-
ment value for recalculation,
or measurement value be-
comes lower than measure-
ment available range after
recalculation.
• Check that measurement value ap-
pears and redo measurement.
• Check that user calibration is cor-
rect and redo measurement.