Konica Minolta CS-200 Camera Accessories User Manual


 
68
Measurement
Preparation
7.
Press either
or
key to select
[USER CAL] and then
ENTER
key.
< USER CAL DATA> screen appears.
8.
Press either
or
key
to select stored data.
Please select data which was measured using CH00
(KONICA MINOLTA SENSING’s calibration standard.)
The data are considered to have been calculated using CH00 (KONICA
MINOLTA SENSING’s calibration standard) without compensation coef
-
ficients for user calibration, even if data measured using a calibration channel
other than CH00 are selected. Because of this, when existing data measured
using a calibration channel other than CH00 are selected for performing
user calibration, user calibration cannot be performed properly.
9.
Press
COLOR
key in SHIFT mode,
and select color space.
Select same color space as calibration value
to be entered in procedures from
11
to
13
.
10.
Press
ENTER
key.
Calibration entry screen appears.
(2) Through selection from saved data
continuation from a p.64, 65.
15 10, 14
12
7, 8, 1199
USER CAL
<
CH SETTING
>
CH01:
<
DEFAULT
>
USER CAL
CH ID NAME
CH COPY
CH RESET
BREAK:[ESC]
[ ][ 2° ]
<USER CAL DATA>
CH01:<DEFAULT>
Lv ------
cd/m
2
x ------
y ------
SELECT MEH DATA
OR MEASURE.
NEXT : [ENTER]
[ ][ 2° ]
<
ANGL
>
1
°
VIEW
<
SPD
>
AUTO
<
SYNC
>
NO SYNC
<
LENS
>
STANDARD
<
MEM
>
AUTO NUM
<MEMORY DATA>
M
002
:
Lv 100.76
x 0.3945
y 0.3998
<
ANGL
>
1
°
VIEW
<
SPD
>
AUTO
<
SYNC
>
NO SYNC
<
LENS
>
STANDARD
<
MEM
>
AUTO NUM
<MEMORY DATA>
M002 :
Lv 100.76
x 0.3945
y 0.3998
<USER CAL DATA>
CH01:<DEFAULT>
Lv ------
cd/m
2
x ------
y ------
SET CAL VALUE.
OK : [ENTER]
[ SFT][ 2° ]
<USER CAL DATA>
CH01:<DEFAULT>
Lv ------
cd/m
2
x ------
y ------
SELECT MEH DATA
OR MEASURE.
NEXT : [ENTER]
[ SFT][ 2° ]
<
ANGL
>
1
°
VIEW
<
SPD
>
AUTO
<
SYNC
>
NO SYNC
<
LENS
>
STANDARD
<
MEM
>
AUTO NUM
<MEMORY DATA>
M002 :
Lv 100.76
x 0.3945
y 0.3998