This chroma meter realizes high-precision measurement of luminance and chromaticity
comparable to spectroradiometers by the employment of newly developed spectral fit-
ting method. Carefully read this manual before using one.
Packaging material
Retain accompanying packaging materials (carton, protector, and plastic bag) and
holding cap (CS-A24) supplied as standard accessory for future usage.
This is delicate measurement instrument. Use packaging materials supplied in pur
-
chasing in case this instrument needs to be transferred for such purpose as main-
tenance in KONICA MINOLTA SENSING's factories. These packaging materials are
useful for minimizing shock or vibration to this instrument in such situation. Use hold-
ing cap for the same purpose especially to protect optical system of this instrument.
Should any of these packaging materials or holding cap (CS-A24) be lost or broken,
please contact the nearest KONICA MINOLTA SENSING authorized service facility.
Note on Use
Operating Environment
Do not use this instrument outdoor since standard accessory AC adapter is designed
for indoor use.
Do not disassemble this instrument for being composed of delicate electronic compo-
nents.
Use this instrument at rated voltage of 100 V - 120 V or 200 V - 240 V .
Connect AC power cord to AC outlet with rated voltage and frequency. Connected
voltage should not be outside the range of +/-10% of nominal.
This instrument is classified into a Pollution Degree 2 as instrument used in mainly
in manufacturing plant, laboratory, warehouse or equivalents. Use this instrument in
metal dust free and non condensing potential environment.
This instrument is categorized into Installation Category II as equipment connected to
commercially available power source.
Connect PC for controlling this instrument to the outlet with protective grounding.
Failure to follow this may result in electric shock due to short circuit.
Take care not to enter foreign substance like water or metal in this instrument. Operat-
ing in such state cause serious danger.
Do not use this instrument under direct sunlight or near heater. The internal tempera-
ture of this instrument to becomes much higher than ambient temperature which may
break this instrument.
Avoid rapid change in ambient temperature which may form dew condensation.
Avoid using this instrument in extremely dusty or humid place.
Use this instrument at ambient temperature between 0 and 40˚C and relative humidity
85% or less (at 35˚C) with no condensation. Operating this instrument outside speci-
fied temperature and humidity range may unsatisfy its original performance.