Konica Minolta CS-200 Camera Accessories User Manual


 
115
Description
Error
No.
Symptom Item to Check Corrective Action
Page to
Refer
9
Display of
the remaining
measurement
time freezes,
and measure
-
ment is not
completed
for the set
measurement
time.
Is object to measure
stable?
When an object whose luminance
has changed greatly from the pre
-
vious measurement is measured,
measurement may be interrupted to
set the optimum gain for measure-
ment and then remeasurement will
be performed; during remeasure-
ment, the remaining measurement
time display is frozen. Keep object to
measure stable.
Are you measuring an
object whose lumi
-
nance has changed
greatly from the
previous measure
-
ment?
When an object whose luminance
has changed greatly from the pre
-
vious measurement is measured,
measurement may be interrupted to
set the optimum gain for measure-
ment and then remeasurement will
be performed; during remeasure-
ment, the remaining measurement
time display is frozen.
10
PC cannot
read output
data from this
instrument
when con
-
nected with
USB. Com
-
mand or data
cannot be
entered from
PC to this
instrument.
Has USB cable been
tightly connected?
Connect this instrument and PC
securely.
98
Hasn't remote mode
been cancelled?
Switch to remote mode by sending
connection command from PC to this
instrument.
Use standard accessory data manage-
ment software CS-S10w Standard.
99
Has prepared program
been correct?
Check referring to sample program. Use
standard accessory data management
software CS-S10w Standard.
-
11
Measurement
data or vari-
ous setting
becomes
invalid soon.
Hasn't backup battery
been insufficient after
a long time period of
nonuse? Has backup
battery been charged
enough at the time of
purchased?
Switch ON to charge backup battery.
Backup battery is to fulfill its func-
tion after fully charged for approx. 20
hours.
Shelf life of backup battery is 10
years under general use, but re-
placement is needed should memo-
ry disappears soon after full charge.
For replacing backup battery, please
contact the nearest KONICA MI-
NOLTA SENSING authorized service
facility.
4
12
Same error
message ap-
pears repeat-
edly.
Check corrective action
for error message.
If symptom does not improve, please
contact the nearest KONICA MI
-
NOLTA SENSING authorized service
facility.
108