7. Settings
-99-
3
Select the Enabled Items.
To add or delete a display item, select the item, and then click or .
4
Set the order in which the items are to be saved.
The items displayed as the Enabled Items will be saved in sequence from the top. To change the
sequence in which the items are saved, select the item, and then click [Up] or [Down].
5
Save the settings and exit the screen.
Click [OK].
7.7.2 Study Logs tab
Select the settings for updating the UIDs if images are to be re-transferred when study logs or past studies
are displayed.
However, these settings do not apply to studies whose transfer status (see page 31) is “untransferred.”
Study Logs tab
Whichever item is selected, only the transfer time is updated.
Furthermore, it is the information of the images to be re-transferred that is updated and not the information
of the images saved.
Item Operation
Update “Study Instance
UID” on re-transfer
If this is checked, the study instance UID is issued as follows.
Model UID. study ID. transfer time (YYYYMMDDHHMMSS)
Update “Series
Instance UID” on re-
transfer
If this is checked, the series instance UID is issued as follows.
Model UID. study ID. transfer time (YYYYMMDDHHMMSS).
series number
Update “SOP Instance
UID” on re-transfer
If this is checked, the SOP instance UID is issued as follows.
Model UID. study ID. transfer time (YYYYMMDDHHMMSS).
series number. instance number (image number)