Canon CR-1 MARK II Camcorder User Manual


 
8. Before calling the service technician (Warning)
-123-
Continued on the following page >>
10049
7
Entered Model No. is not numeric or the value is
out of the range between 1 and 99.
Set the model number to a number from 1 to 99.
8
Length for storage path for the study file is too
long. Shorten to less than 80 bytes.
Set the length of the storage path to less than 80
bytes.
22
Character '\' or control character cannot be used for
Accession No.
The study information in the study file (ccrhis.his) is
invalid. Delete the backslash (\) and any control
characters.
23
Character '\' or control character cannot be used for
Patient ID.
24
Character '\,' line feed code, or control character
cannot be used for Study Description.
The study information in the study file (ccrhis.his) is
invalid. Delete the backslash (\), line feed code, and
any control characters.
25
Character '\', '^' or control code cannot be used for
Patient Name.
The study information in the study file (ccrhis.his) is
invalid. Delete the backslash (\) and any control
characters.
26
Multi-byte character cannot be used for Patient
Name.
The study information in the study file (ccrhis.his) is
invalid. Delete the multi byte characters.
29
Character '\', '^' or control code cannot be used for
Referring Physician.
The study information in the study file (ccrhis.his) is
invalid. Delete the backslash (\) and any control
characters.
39
Out-of-Range Birth Date (A date earlier than 1869
or later than tomorrow is entered.)
The study information in the study file (ccrhis.his) is
invalid. Input the correct birth date.
40 Entered date is invalid for Birth Date.
41
A character/characters for Patient ID is/are not
supported by Specific Character Set.
The study information in the study file (ccrhis.his) is
invalid. Input using the specific character set.
42
A character/characters for Accession No. is/are not
supported by Specific Character Set.
43
A character/characters for Patient Name is/are not
supported by Specific Character Set.
44
Entered characters are not acceptable because they
do not match Specific Character Set.
The study information in the study file (ccrhis.his) is
invalid. Delete the single-byte Japanese kana
characters.
45
A character/characters for Referring Physician is/
are not supported by Specific Character Set.
The study information in the study file (ccrhis.his) is
invalid. Input using the specific character set.
46
A character/characters for Study Description is/are
not supported by Specific Character Set.
52 Folder for study file is not set. Specify the folder to save the study file (ccrhis.his).
10050
2
Length value for Accession No. of Input
Restrictions is not numeric or is not in the range of
1 to 16.
Set the length limitation of the Accession No. to 16
digits or less.
3 Character set for Accession No. is not specified. Specify the input character type.
6
Length value for Patient ID of Input Restriction is
not numeric or is not in the range of 1 to 64.
Set the length limitation of the Patient ID to 64
characters or less.
7 Character set for Patient ID is not specified.
Specify the input character type.
10 Character set for Patient Name is not specified.
11
Character set for Study Description is not
specified.
13 Starting Value is not numeric or out of range.
Set the Starting Value using a number in the correct
range.
15
Total length of Starting Value and Accession No.
Header String exceeds the limit.
Adjust the length of the header string to set the total
length of the starting value and header string within
the length limitation.
16 Counter value is set to 1. Check the value of the counter.
17
Header String for Patient ID exceeds the
maximum length.
Set the length of the Header string within the length
limitation.
module id message id Message Detail