8. Before calling the service technician (Warning)
-118-
10011
104 Enter User ID/Password. When logging in, input the user ID and password.
105 Enter in alphanumeric.
Input the user ID/password in single-byte
alphanumeric characters.
109
Application will be closed.
Press Cancel button and log in to the application to
view the details of the transfer error.
To check the details of the transfer error, click
[Cancel], and then log in.
10012 28
Image can be captured but the storage space for
captured images runs short.
Make up space in the drive by deleting studies in
Study Logs List.
Delete studies in the study log screen to increase the
amount of available space in the drive.
10013
50
Images stored for more than the specified periods
will be deleted.
To delete the image, click [OK].
66
All the studies will be closed and the current user
will be logged off.
To log off, click [OK].
85 Application cannot be terminated during transfer.
Close all studies after transfer, then exit the
software.
86
All the studies will be closed and then the
application will be.
To end the study and exit the software, click [OK].
10019
10
A study with same Accession No. or Patient ID is
now in progress.
Do you want to create a new study ?
To start as a new study, click [YES].
12
A study with same Accession No. or Patient ID is
now in progress.
Do you want to update ?
To update the study information, click [YES].
10020
1000
No Storage Selected
Image(s) will not be transferred.
Select the storage, and then transfer the image.
1001
Transfer for this study is in progress.
You can't transfer it before the transfer is
completed.
Wait until the transfer is completed before
performing the transfer.
10023
1 Invalid Values for Birth Date Input the correct birth date.
5
Patient Info. set as any of essential data items are
not entered.
Input the required item that is displayed.
6
The length of either of 5 Patient Name fields is too
long.
Shorten the length to 60 characters or less.
Shorten the length to 60 characters or less.
12 Invalid character was entered to the field.
Re-input the data based on the information displayed
in [Details].
14
Line feed code cannot be entered to Study
Description field.
Delete the line feed code.
15
Too Long Patient Name.
Shorten the total length of the 5 items to 60
characters or less.
Input the Patient name in 60 characters or less for all
five items total.
16 Length of Accession No. exceeds the limit. Input within the limited length.
17
A character/characters for Accession No. is/are not
supported by Acceptable Character Set or Specific
Character Set.
Input using the designated input character type or
specific character set.
18 Length of Patient ID exceeds the limit. Input within the limited length.
19
A character/characters for Patient ID is/are not
supported by Acceptable Character Set or Specific
Character Set.
Input using the designated input character type or
specific character set.
21
A character/characters for Study Description is/are
not supported by Acceptable Character Set or
Specific Character Set.
22
Character '\' or control character cannot be used for
Accession No.
Delete the backslash (\) and any control characters.
23
Character '\' or control character cannot be used for
Patient ID.
module id message id Message Detail