8. Before calling the service technician (Warning)
-124-
10050
18
Character '\' or control character cannot be used for
Patient ID.
Delete the backslash (\) and any control characters.
19
A character/characters for Patient ID is/are not
supported by Acceptable Character Set or Specific
Character Set.
Input using the designated input character type or
specific character set.
20
Header String for Accession No. exceeds the
maximum length.
Set the length of the Header string within the length
limitation.
21
Character '\' or control character cannot be used for
Accession No..
Delete the backslash (\) and any control characters.
22
A character/characters for Accession No. is/are not
supported by Acceptable Character Set or Specific
Character Set.
Enter using the designated input character type or
specific character set.
23
Header String for Study Description exceeds the
maximum length.
Set the length of the Header string within the length
limitation.
24
Character '\' or control character cannot be used for
Study Description.
Delete the backslash (\) and any control characters.
25
A character/characters for Study Description is/are
not supported by Acceptable Character Set or
Specific Character Set.
Enter using the designated input character type or
specific character set.
26 Patient/Study Info. Setting Error
Set again the Patient/Study Info. based on the
information displayed in [Details].
29
Reset the maximum Length and Header string for
Accession No. so that Counter value will be two
digits or more.
Refer to the message details, and set the Accession
No.
10053 1000
This type of a magnetic card/bar code reader
cannot use Connection Check function.
Check the magnetic card/barcode reader settings.
10063
1001
Enter essential data.
[item]
Input the required item that is displayed.
1002
Enter in numeric.
[item]
Input the displayed item in single-byte alphanumeric
characters.
1003
Entered value exceeds the limit.
[item (*** - ***)]
Input the displayed item within the limitation.
1004 Entered value could not be checked.
Check the input value in the JPEG Settings
screen (see page 70), and set it again if necessary.
1005
Entered value is not numeric or out of range.
[item]
Enter a value within the correct range for the
displayed item.
10065 4097 C/D Ratio data will be updated.
Click [OK] to overwrite the C/D ratio data.
To cancel overwriting, click [Cancel].
module id message id Message Detail