Using the Service and Assembly Module (SAM)
May 2002 A-5
Performing the Functions on the Mag/Focus Tab
IMPORTANT: Magnification Calibration should be done only with the guidance of
a Kodak support representative.
From the Operations window, click the Mag/Focus tab when you need to set the
magnification (position of the lens) or do an autofocus.
Setting the Magnification
1. On the Mag/Focus tab, select the Chosen Magnification. The value can be
between 0.5 and 2.0.
The scanner’s software maintains an internal Mag/Focus position table that
associates a motor position for each of the possible magnification values.
2. Click Do Magnification.
The lens and CCD move to the chosen position.