B&K 8510 Digital Camera User Manual


 
34
15) Repeat 9
)~14),
and set the other process step by step as follows:
A.
CONST CURRENT,0A,SHORT OFF, READ BACK V, 5.9V, 6.4V,1S
B.
CONST VOLTAGE, 5V, SHORT OFF,READ BACK A, 0.205A, 0.245A,1S
C.
CONST VOLTAGE, 3V, SHORT OFF, READ BACK A, 0.205A,0.245A,1S
D.
CONST VOLTAGE, 2V,SHORT OFF, READ BACK A, 0.205A,0.245A, 1S
E.
CONST CURRENT, 0A, SHORT ON, READ BACK A,0A,0.245A, 1S
16) VFD displays
SHORT TEST FILE*,
which requests saving the files edited to
EEPROM,
automatically testing
files and
LIST
files use a common memory area, and the area can contain 8 groups of automatic testing files for
next usage. Here, the testing file could be saved in the first group, press
1
and press
Enter
to confirm.
17) When the edition of automatic testing file is finished, press
Enter
two times to back the menu.
Taking out the Testing File Rapidly
1) Press
Shift
+
Menu
to enter into menu, VFD displays
>
>>
>CONFIG
2) Pres
s
and move the menu to
>LIST SET
3) Press
Enter
to enter into submenu, VFD displays
>MODE SET
4) Press
and move the menu to
>CALL TEST FILE
5) Press
Enter
and recall the file edited
Automatic Testing
After editing the automatic testing file, the automatic testing could be carried out by the below steps.
1) Press
Shift
+
I-setI-set
to enter into automatic testing, VFD displays the name of current file:
NAME: TEST
FILE1
2) Press
Enter
to observe the voltage and current we inputted, and also can observe the step number of testing.
3) Press
Shift
+
Trigger
and begin to operate, when the automatic testing is pause (the current prolong time of
one step is 25.5S), press
Shift
+
Trigger
also can make the testing continue.
4) When the automatic testing is finished, the buzzer will hint by tweet (buzzer), VFD reads the result of testing. If the
testing passed, VFD reads PASS, or reads FAULT. Pres
s▲、
also to observe the numerical value and result
of every step.
5) Press
ESC
to back the automatic testing.
Setting the Resolution and number of decimal points
By selecting the low range and high range, you can measure the voltage and current more precisely.