40. Calibrate current value
64H
1
st
byte Start bit ( AAH )
2
nd
byte Address (0—0XFE)
3
rd
byte
Command
64H
4
th
byte
Current calibration point
1~4
From 5
th
to 25
th
byte System reserve
26
th
byte Sum code
NOTE
Current calibration standard points have four: 1,2,3,4
41. Sending the actual input current to calibration program (65H
1
st
byte Start bit ( AAH )
2
nd
byte Address (0—0XFE)
3
rd
byte
Command
65H
4
th
byte The lowest byte of actual current
5
th
byte The lower byte of actual current
6
th
byte The higher byte of actual current
7
th
byte The highest byte of actual current
From 8
th
to 25
th
byte System reserve
26
th
byte Sum code
42. Store the calibration data to EEPROM
66H
1
st
byte Start bit ( AAH )
2
nd
byte Address (0—0XFE)
3
rd
byte
Command
66H
From 4
th
to 25
th
byte System reserve
26
th
byte Sum code
NOTE
Finish the calibration operation, users should save the calibration parameter in EEPROM with this command; users
could use these data in next power on.
43. Setting / Reading calibration information
67H/68H
1
st
byte Start bit ( AAH )
2
nd
byte Address (0—0XFE)