Tektronix P6330 Digital Camera User Manual


 
Operating Basics
P6330 3.5 GHz Differential Probe Instructi on Manual
15
Probe Grounding
In addition to the plus and minus inputs on the probe head, there is
also a ground (common) input. The ground lead slides into the notch
on the side of the probe. See Figure 7.
+
--
Ground
Figure 7: Probe ground input
CAUTION. To avoid damaging the circuitry under test, connect the
probe ground (common), if used, to a ground-reference point only.
In most applications, the common-mode impe dance to ground is
greater than the differential impedance. Adding the probe ground
lead does not improve the high-frequency performance of the
measurement. You can use the probe to take a differential measure-
ment regardless of whether or not the ground (common) is con-
nected.
There are some applications that may require a ground reference
connection to maintain measurement accuracy. Generally this is
necessary when probing circuits which are fully isolated from
ground, such as battery operated devices.