Tektronix P6330 Digital Camera User Manual


 
Performance Verification
P6330 3.5 GHz Differential Probe Instructi on Manual
39
Probe Calibration Fixture
Some of the procedures in this manual use a probe calibration
fixture, Tektronix part number 067-0419-00.
The calibration fixture provides a means to test the probe for bot h
common mode and differential mode measurements. SMA
connectors allow stimulus signals to connect to the fixtu re an d are
located on the front and back of the fixture. The fixt ure i s designed
to be used with a probe positioner, such as a Tektronix PPM203B.
Figure 16: Probe Cal ibration Fixture
Using the Probe Calibration Fixture
1. Connect the fixture to the test circuit using an SMA cable.
2. Connect the 50 terminator included with the fixture to the
unused SMA connector.
3. Insert and secure the probe in a probe positioner.
4. Position the probe over the fixture, using either the positioner
coarse adjustment or otherwise manipulating the positi oner arm
in place.