Tektronix P6330 Digital Camera User Manual


 
Performance Verification
44
P6330 3.5 GHz Differential Probe Instruction Manual
If the measurement reading is not stable, use Average mode
(16 Averages) to improve stability.
The system rise time (t
s
) that you measured in step 7 represents the
rise time of the test system without the probe. The system rise time is
used to calculate the probe rise time (t
p
)instep16.
The following steps instruct you to assemble the test setup that
includes the probe, as shown in Figure 21. The system and probe rise
time (t
s+p
)thatyoumeasureinstep15isusedtocalculatetheprobe
rise time (t
p
)instep16.
Oscilloscope
P6330
1103
CH 1
input
CH 1
output
BNC-to-SMA
adapter
015-0572-00
SMA cable
015-0562-00
SD24/ 80E04
CH1
CH2
Probe cal fixture
067-0419-00
SMA cable
174-1427-00
Figure 21: Test system rise time setup with probe