Performance Verification
44
P6330 3.5 GHz Differential Probe Instruction Manual
If the measurement reading is not stable, use Average mode
(16 Averages) to improve stability.
The system rise time (t
s
) that you measured in step 7 represents the
rise time of the test system without the probe. The system rise time is
used to calculate the probe rise time (t
p
)instep16.
The following steps instruct you to assemble the test setup that
includes the probe, as shown in Figure 21. The system and probe rise
time (t
s+p
)thatyoumeasureinstep15isusedtocalculatetheprobe
rise time (t
p
)instep16.
Oscilloscope
P6330
1103
CH 1
input
CH 1
output
BNC-to-SMA
adapter
015-0572-00
SMA cable
015-0562-00
SD24/ 80E04
CH1
CH2
Probe cal fixture
067-0419-00
SMA cable
174-1427-00
Figure 21: Test system rise time setup with probe