Tektronix P6330 Digital Camera User Manual


 
Performance Verification
38
P6330 3.5 GHz Differential Probe Instruction Manual
Table 7: Equipment required for performance verification (Cont.)
Item description Recommended examplePerformance requirement
Precision coaxial cable Male-to-Male SMA, 1 ns
delay
015-0562-00
Adapter SMA Female-to-Female 015-1012-00
Adapters (2) SMA Female-to-BNC Male 015-0572-00
Adapter SMA Male jack 131-7128-00
Adapter BNC Female-to-Dual
Banana
103-0090-00
Adapter Square pin adapter 016-1884-00
Adapter Y-lead adapter 196-3434-00
Adapters (2) KlipChip adapter 206-0364-00
Equipment Setup
Use this procedure to set up the equipment to test the probe.
1. Connect the probe to the 1103 power supply.
2. Turn on 1103 power supply.
3. Turn on the oscilloscope.
4. Allow 20 minutes for the equipment to warm up.