P6330 3.5 GHz Differential Probe Instructi on Manual
47
Adjustments
The P6330 has 3 internal controls: offset zero, offset range, and
DC CMRR. These controls should only be adjusted after a probe
performance verification and functional check has been performed
on the oscilloscope, and only if a check fails to meet its specifica-
tion.
To make adjustments to the probe, the compensation box cover must
be removed and the equipment allowed to warm up for 20 minutes.
Equipment Required
In addition to the equipment required to perform the performance
verification, the adjustment procedures require the equipment listed
in Table 8.
Table 8: Additional equipment required for adjustment
Item description Recommended example
Release tool 003-1383-00
Adjustment tool 003-0675-01