41
Shooting mode Normal / SP / Stereo (electrically switched)
Stereoscopic photography method Simultaneous stereo photography
Stereo photography parallax 7.4° (at the 0 diopter eye)
Field angle
Normal mode : 45°
SP mode : 45°
Stereo mode : 34° (20° horizontal and 27° vertical directions)
*
Some eyes may cause a are around their circumference.
Tolerance of pixel pitch on fundus
(ISO 10940)
Normal image : S
*
( 9 μm ± 7 %)
L
*
(4.5 μm ± 7 %)
Stereo image : S
*
(5.5 μm ± 7 %)
L
*
(2.7 μm ± 7 %)
Working distance 30 mm
Minimum pupil size
Normal mode :
4.0 mm
SP mode :
3.5 mm
Stereo mode :
4.0 mm
Compensation range of
examined eye
Withou
t compensation : −12 m
-1
(D) to +13 m
-1
(D)
Compensation + : +10 m
-1
(D) to +35 m
-1
(D)
Compensation − : −32 m
-1
(D) to −10 m
-1
(D)
Focusing Split luminous bars coincidence
Working distance adjustment 2 luminous dots indication type
Camera Specic Nikon digital SLR camera
Monitor 5.7-inch LCD monitor
Light source
Observation
: Near-infrared LED
Photography : Xenon ash lamp
Internal xation target Central, Disc, Macula,mosaic 8 positions
Opti
cal head base adjustment range
Movable 40mm forward/backward
Movable 98mm leftward/rightward
Movable 27mm vertically (electric)
Chin rest adjustment range Movable 55mm (electric)
Interfaces
Image output : USB (B type)
Imaging conditions output
: USB (mini-B type )
Numerical keypad input : USB (A type)
Power supplies
Input : AC100-240V, 50/60Hz
Power consumption : 150V
A
Dimensions 310(W)x504(D)x548(H)mm
Weight 21kg / 46lbs (excluding the attached digital SLR camera)
*S/L : Image size set on the digital camera
Compliance standard
EN 60601-1:1990+A1:1993+A2:1995
EN 60601-1-2:2001+A1:2006
Classication of equipment based on EN 60601-1
According to the type of protection against electric shock
(Class
device)
According to the degree of protection against electric shock
(Type B applied part)
According to the type of protection against ingress of water as detailed in the current edition of IEC60529.
(IPX0)
According to the degree of safety of application in the presence of a ammable anaesthetic mixture with air or with
oxygen or nitrous oxide.
(Equip
ment not suitable for use in the presence of a flammable anesthetic mixture with air or with
oxygen or nitrous oxide)
According to the mode of operation.
(continuous operation)
Complied EC Directive
Medical Device Detective 93/42/EEC :
WEEE Directive 2002/96/EC