Kowa nonmyd WX 3D Digital Camera User Manual


 
44
12.2 digital camera setting / 12.3 Imaging condition output
12.3 Imaging conditions output
This instrument outputs the imaging conditions from Imaging condition output terminal. If you want to read the imaging
conditions using application software other than that provided with the instrument, refer to the specications below:
For this purpose, you need to install FTDI
“Direct (D2XX) driver Ver 2.04.06 or later”
into your PC to which this instru-
ment is connected.
Communication related specications
Baud rate : 38400
Data length : 8 bit
Parity : even
Stop bit : 1 bit
Flow control : none
Output data
Output item Description Example
Model name Fixed value K9L57
Instrument number 10-digit number 16174
]]]]]
Photography mode NORMAL/SP/STEREO STEREO
Light intensity selector knob
position
-2/-1/0/+1/+2 0
Blue or brown eye BLUE/BROWN BLUE
Diopter compensation 0/D+/D- 0
Left or right eye L/R R
Fixation target 1 to 7/EXT 1 to 7
Internal xation target illuminat-
ing position
C (Posterior), N (optic disc), M (Macula), U (Up)
UR (Upper right), UL (Upper left), R (Right), L (Left), D (Down),
LR (Lower right), LL (Lower left)
C
Observation screen ANTE/FUNDS FUNDS
k
A linefeed code is inserted every output of an item.