Operator’s Manual LBA-PC
256
0 = reference frame
NOTE: The detector origin is always the upper left corner so that y values increase going down and decrease going
up.
Returns:
PNW <configuration>
2
Key Type Value Description
UpperLeft I,I X,Y - minimum allowable capture location
LowerRight I,I X,Y - maximum allowable capture location.
Note that the capture location plus the
capture size times the capture resolution
must be less than these values.
A.5.4.13 PFS? - read pass/fail status
:PFS?
Returns:
PFS <configuration>
Note: Only results that are tested are returned.
Key Type Value Description
Label S Label string as displayed in the Results
window.
Status B 0 = fail,
1 = pass
Example return:
:PFS Total=1;Peak=0;Centroid X=0;Centroid Y=1;;
Total, Peak, Centroid X, and Centroid Y are being tested. Total passed, Peak failed, Centroid X
failed, Centroid Y passed.
A.5.4.14 RCC? - read cursor column
:RCC? <configuration>
Key Type Value Description
FrameNumber I Frame number
-1 = gain frame