Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-1
Model MicroCEM
SECTION 3
OPERATION
CAUTION.
Do not operate or service this instru-
ment before reading and understanding
this instruction manual and receiving
appropriate training.
Refer to installation drawing supplied
with the application data package.
3-1 STARTUP PROCEDURE
Once the MicroCEM has been correctly as-
sembled and installed in accordance with
the instructions in Section 2, the analyzer is
ready for operation.
Before operating the system, verify that the
leak checks have been performed and that
the sample handling unit is performing cor-
rectly.
Apply power to the system and verify that
sample gas is flowing.
NOTE
A warm-up time of from 15 to 50 minutes
is required depending on the installed
detector(s).
Analyzer operation can be confirmed on the
screen of the pocket PC through the glass
window on the door. Upon power up, the
analyzer will perform a self-test routine.
This test will last approximately 60 sec-
onds.
3-2 POCKET PC USER INTERFACE
The MicroCEM User Interface runs on a
Pocket-PC with Windows CE operating
system. It communicates with the Micro-
CEM via serial communication port. All in-
put to the Pocket-PC is done using a
pointing device that comes with the Pocket-
PC.
a. Connecting Pocket PC to MicroCEM
1. Open MicroCEM door. Refer to
Figure 3-2.
2. Plug RS232 plug into adapter lo-
cated on front panel.
3. Plug power supply cable into 5V
adapter
4. Turn Pocket PC on
5. In order to assure no other win-
dows are open press the reset
button. Reset button is located on
the back of the pocket PC.
6. Using the MicroCEM menu click on
“programs”
7. Click on “Connection” icon
8. Click on “uCEM” icon.
9. Go to tools menu and click on Mi-
croCEM .
10. Unit will display data in 3 to 5 sec-
onds.
Figure 3-1. Pocket PC